Ultra Deluxe Cold Rod Insert™
Model ECT/UL-DLX/I
Main Features:
-Designed for use with the 21.6 cm inside diameter flanged phantoms
Main Applications:
-For use with ultra-high spatial resolution SPECT
and PET Systems
-System resolution test for lower contrast structures
-Acceptance testing
-Routine quality, assurance and control
-System performance evaluation
-Research
Specifications:
-Insert diameter: 21.3 cm
-Rod diameters: 3.2, 4.8, 6.4, 7.9, 9.5 and 11.1mm
Deluxe Cold Rod Insert™
Model ECT/DLX/I
Main Features:
-Designed for use with the 21.6 cm inside diameter flanged phantoms
Main Applications:
-For use with high and very high spatial resolution SPECT and PET Systems
-System resolution test for lower contrast structures
Acceptance testing
-Routine quality, assurance and control
-System performance evaluation
-Research
Specifications:
Insert diameter: 21.3 cm
Rod diameters: 4.8, 6.4, 7.9, 9.5, 11.1 and 12.7 mm
Height of rods: 8.8 cm
Standard & Elliptical Cold Rod Insert™
Model ECT/STD/I and Model ECT/ELP/I
Main Features:
-Model ECT/STD/I is designed for use with the 21.6 cm inside diameter flanged phantoms
-Model ECT/ELP/I is designed for use with the Elliptical ECT Phantom™ (Model ECT/ELP/P)
Main Applications:
-For use with medium to high spatial resolution SPECT and PET Systems
-System resolution test for lower contrast structures
-Acceptance testing
-Routine quality, assurance and control
-System performance evaluation
-Research
Specifications:
Insert diameter:
Model ECT/STD/I: 21.3 cm
Model ECT/ELP/I: 30.2 cm along major axis
21.8 cm along minor axis
Rod diameters: 6.4, 7.9, 9.5, 11.1 12,7 and 19.1 mm
Height of rods: 8.8 cm
Benchmark Cold Rod Insert™
Model ECT/BEN/I
Main Features:
-Designed for use with the 21.6 cm inside
diameter flanged phantoms
Main Applications:
-For use with ECT systems with poorer spatial resolution, larger radius-of-rotation, cut-off reconstruction filters and count densities mimicking clinical scans
-System resolution test for lower contrast structures
-Acceptance testing
-Routine quality, assurance and control
-System performance evaluation
-Research
Specifications:
Insert diameter: 21.3 cm
Rod diameters: 9.5, 11.1, 12.7,15.9. 19.1, and 25.4 mm
Height of rods: 8.8 cm
|